Principle - Bead Injection
BI exploits flow programming to meter, transport, capture , perfuse, monitor and discharge microspheres in order to separate and assay species that can be bound and released from immobilized functional groups. BI uses microliter volumes of Sepharose or other chromatographic materials available with a wide variety of functional groups. The BI technique can be configured in two ways: as a microSI chromatography or as BI spectroscopy.
BI opens unexplored avenues for development of novel (bio)analytical assays, for research in solid-liquid interactions, for study of quality of chromatographic supports, and for optimal of immobilization of bioligands on solid supports. BI is far more versatile and less expensive than analytical methods based on the use magnetic beads.
microSI Chromatography on Renewable Column
The assay protocol for microSIC has the following steps :
A) A precise volume of bead suspension is aspirated via MPV, and by flow reversal dispensed downstream. where the beads are captured in a well defined geometry in a close proximity to the flow cell (see insert).
B) The MPV is switched towards the sample source, a precise volume of sample solution is aspirated, and after MPV is switched towards the column, the flow is reversed .
C) The injected sample zone passes through the column where the analytes are retained.
D) The matrix as well as any not retained components are washed out.
E) A precise volume of eluent is aspirated, the MPV is switched and the eluent is directed trough the column. The eluted analyte is monitored as it passes through the flow cell
F) The microcolumn is discarded.
Example: microAffinity Chromatography
Bead Injection Spectroscopy
The assay protocol for BI Spectroscopy has the following steps :
A) A precise volume of bead suspension is aspirated via MPV, and by flow reversal dispensed downstream, where the beads are captured in a well defined geometry within the flow cell (see insert).
B) The MPV is switched towards the sample source, a precise volume of sample solution is aspirated, and after the MPV is switched towards the flow cell and the flow is reversed .
C) The injected sample zone passes through the flow cell, where the analytes are retained on bead surfaces and detected. Note (insert), that the response curve has a plateau, if the retained analyte is bound firmly to bead surfaces.
D) The matrix, not retained components is washed out.
E) A small volume of eluent is aspirated, the MPV is switched and the eluent is directed trough the flow cell. If all retained species are eluted the signal will reach baseline. If not the beads within the flow cell can be automatically removed.
F) The beads are discarded from the flowcell.
Example: BI Spectrosopy of Native Insulin
Sephadex and Sepharose beads are ideal for BI applications since they are transparent allowing target molecules to be monitored by UV-VIS and by fluorescence. These beads are available with a wide variety of functional groups: cation and anion exchange , C-18, Protein A, Protein G , avidin and streptavidin. They vary in diameter between 30 and 150 micrometers.